Reports until 00:41, Friday 10 March 2017
H1 General
nutsinee.kijbunchoo@LIGO.ORG - posted 00:41, Friday 10 March 2017 - last comment - 10:29, Friday 10 March 2017(34716)
Ops Owl Shift Transition

Relocking is going okay. Stopping at Violin_Mode_Damping_2 to squish one of the fundamental modes that rung up high.

Comments related to this report
nutsinee.kijbunchoo@LIGO.ORG - 00:45, Friday 10 March 2017 (34717)

A bunch of ETMX CDS overview went red briefly, middle ish column (either DAC or DAQ). Came with OMC DC PD saturation warning. And the red went away before I could look closer.

nutsinee.kijbunchoo@LIGO.ORG - 00:56, Friday 10 March 2017 (34718)

Accepted ETM and ITM SDF differences after making sure ETMX bounce mode was not ringing up. The rest are gain settings. Back to Observe.

Images attached to this comment
nutsinee.kijbunchoo@LIGO.ORG - 01:02, Friday 10 March 2017 (34719)

An accepted PI differences. Had to changes mode 28 filter and PLL set freq. Back to Observe again.

Images attached to this comment
david.barker@LIGO.ORG - 10:29, Friday 10 March 2017 (34726)

Nutsinee is correct, at the same time we got a SUS-ETMX glitch the OMC PD DC ADC channels saturated. The time was 08:42:49 UTC. The saturations stopped 2 seconds later, the glitch was cleared at 08:44:04 UTC (cronjob running on the minute). Could this be a wierd coincidence?