WP6522
During the maintenance period yesterday, I have measured the spectra of IMC WFS DCs with and without the IMC locked. The purpose of this measurement was to cross-check the calibration factor of 1/3 which previously estimated by Daniel (31631, Nov 2016).
A conclusion:
I agree with Daniel that the IMC WFS DC signals should be scaled by a factor of 1/3 when the IMC is locked.
[Some plots]
A side note: depending on whether the IMC is locked or not, the spectral shapes vary. This is not so surprising because when the IMC is locked, the reflection port must be dominated by non-00 modes. So the signals should contain relatively large amount of cross terms from various HOMs. So for the reason, I don't have a good explanation why they have to be a factor of 1/3. Even so, it is interesting to see some peaks come and go.