PSL Weekly Report - FAMIS #7431

Laser Status:
SysStat is good
Front End Power is 34.09W (should be around 30 W)
HPO Output Power is 165.3W
Front End Watch is GREEN
HPO Watch is GREEN

PMC:
It has been locked 0 days, 6 hr 15 minutes (should be days/weeks)
Reflected power = 16.57Watts
Transmitted power = 60.79Watts
PowerSum = 77.36Watts.

FSS:
It has been locked for 0 days 4 hr and 17 min (should be days/weeks)
TPD[V] = 3.368V (min 0.9V)

ISS:
The diffracted power is around 3.7% (should be 3-5%)
Last saturation event was 0 days 4 hours and 45 minutes ago (should be days/weeks)

Possible Issues:

Xtal Chiller "light" is showing intermittent alarm on Statues Screen. 300ml of water was added