Reports until 23:52, Wednesday 29 March 2017
H1 SEI
patrick.thomas@LIGO.ORG - posted 23:52, Wednesday 29 March 2017 - last comment - 09:08, Thursday 30 March 2017(35209)
H1 ISI CPS Sensor Noise Spectra Check - Weekly
FAMIS #6891

H1:ISI-ITMX_ST1_CPSINF_V1_IN1_DQ is definitely elevated.
Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 09:08, Thursday 30 March 2017 (35219)

Thanks Patrick.  Yes, this is elevated.  It isn't extreme but it is certainly not like the others.  Better we catch it here than wait for it to get bad and be noisy in band.  We'll start with a gauge board un seat/reseat maneuver next opportunity.

FRS 7763.