Based on the last two stretches of the BRSY DRIFTMON, this signal is trending down at 180 cts/day, see attached. It is currently sitting at -13500cts and it looks like this is a pretty good spot in the disturb/recover cycle for forecasting.
Next Tuesday 4 April, it should be at ~-14500cts, that is okay for operation. If the BRS is not recentered until 11 April, the DRIFTMON will be ~-15700cts. Not sure if we can wait that long. Krishna is contemplating coming over for the 4 April Maintenance day for this.