Displaying report 1-1 of 1.
Reports until 15:46, Friday 20 July 2012
X1 SUS
james.batch@LIGO.ORG - posted 15:46, Friday 20 July 2012 (3531)
Restart X1 triple test stand
Found x1tripleteststand had time frozen at Wednesday evening (July 18).  Could not get models to run, even after restarting the tripleteststand computer and cycling power with the I/O chassis power switch.  Appeared the DAC and ADC cards could not be found.

Ended up turning off power to computer, then turning off power supply to I/O chassis, then powering back up in order.  Started IOP model, then user model, both started properly.

Given the timing, possibly a lightning induced power glitch?  

Displaying report 1-1 of 1.