Reports until 11:12, Tuesday 04 April 2017
H1 SEI (SEI)
krishna.venkateswara@LIGO.ORG - posted 11:12, Tuesday 04 April 2017 (35313)
BRS-Y Recentered

Krishna

I recentered the BRS-Y this morning. The driftmon should end up around + 5k counts, which gives ~20k counts of useful range before the next recentering. The drift rate currently is about 180 counts per day (see image), so next recentering will be in ~3 months. As noted in 34145, recentering was much easier this time because the recentering rod prefers certain positions, which happened to be in an acceptable range this time.

BRS-Y is ready for SEI use.

Images attached to this report