Reports until 02:05, Wednesday 05 April 2017
H1 SEI (SEI)
cheryl.vorvick@LIGO.ORG - posted 02:05, Wednesday 05 April 2017 (35333)
SEI ground seismometer mass position check - Monthly

STS CENTERING: 2017-04-05 02:03:39.354195


All STSs prrof masses that within healthy range (< 2.0 [V]). Great!

Here's a list of how they're doing just in case you care:
STS A DOF X/U = -0.501 [V]
STS A DOF Y/V = 0.243 [V]
STS A DOF Z/W = -0.639 [V]
STS B DOF X/U = 0.524 [V]
STS B DOF Y/V = 0.316 [V]
STS B DOF Z/W = -0.297 [V]
STS C DOF X/U = 0.361 [V]
STS C DOF Y/V = 0.847 [V]
STS C DOF Z/W = -0.281 [V]
STS EX DOF X/U = 0.081 [V]
STS EX DOF Y/V = 0.613 [V]
STS EX DOF Z/W = 0.066 [V]
STS EY DOF X/U = 0.097 [V]
STS EY DOF Y/V = 0.103 [V]
STS EY DOF Z/W = 0.463 [V]

T240 CENTERING:  2017-04-05 01:56:53.355462


There are 6 T240 proof masses out of range ( > 0.3 [V] )!
ETMX T240 2 DOF X/U = -0.615 [V]
ETMX T240 2 DOF Y/V = -0.708 [V]
ETMX T240 2 DOF Z/W = -0.341 [V]
ETMY T240 3 DOF Z/W = 0.312 [V]
ITMX T240 1 DOF X/U = -0.423 [V]
ITMX T240 3 DOF X/U = -0.385 [V]


All other proof masses are within range ( < 0.3 [V] ):
ETMX T240 1 DOF X/U = 0.057 [V]
ETMX T240 1 DOF Y/V = 0.069 [V]
ETMX T240 1 DOF Z/W = 0.143 [V]
ETMX T240 3 DOF X/U = 0.079 [V]
ETMX T240 3 DOF Y/V = -0.011 [V]
ETMX T240 3 DOF Z/W = 0.052 [V]
ETMY T240 1 DOF X/U = 0.012 [V]
ETMY T240 1 DOF Y/V = -0.02 [V]
ETMY T240 1 DOF Z/W = -0.189 [V]
ETMY T240 2 DOF X/U = 0.18 [V]
ETMY T240 2 DOF Y/V = -0.205 [V]
ETMY T240 2 DOF Z/W = 0.037 [V]
ETMY T240 3 DOF X/U = -0.206 [V]
ETMY T240 3 DOF Y/V = 0.005 [V]
ITMX T240 1 DOF Y/V = -0.167 [V]
ITMX T240 1 DOF Z/W = -0.103 [V]
ITMX T240 2 DOF X/U = -0.11 [V]
ITMX T240 2 DOF Y/V = -0.12 [V]
ITMX T240 2 DOF Z/W = -0.141 [V]
ITMX T240 3 DOF Y/V = -0.101 [V]
ITMX T240 3 DOF Z/W = -0.042 [V]
ITMY T240 1 DOF X/U = 0.024 [V]
ITMY T240 1 DOF Y/V = -0.003 [V]
ITMY T240 1 DOF Z/W = 0.007 [V]
ITMY T240 2 DOF X/U = 0.148 [V]
ITMY T240 2 DOF Y/V = 0.168 [V]
ITMY T240 2 DOF Z/W = 0.063 [V]
ITMY T240 3 DOF X/U = -0.067 [V]
ITMY T240 3 DOF Y/V = 0.091 [V]
ITMY T240 3 DOF Z/W = -0.012 [V]
BS T240 1 DOF X/U = -0.13 [V]
BS T240 1 DOF Y/V = 0.007 [V]
BS T240 1 DOF Z/W = 0.264 [V]
BS T240 2 DOF X/U = 0.109 [V]
BS T240 2 DOF Y/V = 0.241 [V]
BS T240 2 DOF Z/W = 0.031 [V]
BS T240 3 DOF X/U = 0.081 [V]
BS T240 3 DOF Y/V = -0.055 [V]
BS T240 3 DOF Z/W = -0.062 [V]