H1 SUS (ISC, SUS)
sheila.dwyer@LIGO.ORG - posted 16:40, Thursday 06 April 2017 - last comment - 22:52, Thursday 06 April 2017(35371)
Op Lev damping off on all test masses, a look at top mas damping

Brett, Jeff and I have been looking a little at Brett's suspension model, and trying to asses how our damping design (both top mass and oplev) is at high power, based on some conversations at the LLO commissioning meeting.  Today I turned off oplev damping on all test masses. 

Top mass damping:

The first attached plot shows the magnitudes of the top to top transfer functions for L, P, and Y for 3 different states (no damping and no power in the arms, top mass only damping and no power in the arms, and top mass only damping and 30 Watts of input power).  You can see that while it would be possible to design better damping loops to deal with the radiation pressure changes in the plant, there isn't anything terrible happening at 30 Watts.  I also looked at 50 Watts input power, and these aren't any large undamped features.  I'm not plotting T, V, and R because the radiation pressure doesn't have any impact on them according to the model.  

Oplev damping:

The next three plots show the transfer functions from pum to test mass pitch, in the optic basis, hard and soft.  One thing that is clear is that we should not try to damp the modes above 1 Hz in the optic basis, so if we really want to use oplev damping we should do it in the radiation pressure basis and probably have a power dependent damping filter.  The next two plots show the impact on the hard and soft modes. You can see that the impact of the oplev damping on the hard modes (and therefore the hard ASC loops) is minimal, but there is an impact on the soft loops around half a Hz.  DSOFT pitch has a low bandwidth, but we use a bandwidth of something like 0.7Hz for CSOFT P to suppress the radiation pressure instability.  

Krishna had noted that the ITMY oplev seems to have been causing several problems recently, (aog 34351), and this afternoon we think that this caused us a lockloss.  I tried just turning the opelv damping offf, which was fine, except that we saw signs of the radiation pressure instability returning after a few minutes at 30 Watts.  I turned the gain up by a factor of 2 (from 0.2 to 0.4), and this instability went away.  The last attached plot is a measurement of the csoft pitch loop with 2 Watts, no oplev damping, before I increased the gain to 0.4.  

The guardian is now set to not turn on the oplev damping, and this is accepted in SDF.  Hopefully this saves us some problems, but it is probably still a good idea to fix the ITMY oplev on tuesday. 

Images attached to this report
Non-image files attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 21:12, Thursday 06 April 2017 (35374)ISC

I don't know if we can run like this. 5 hours in and all of the ASC pitch loops are angry and the range is suffering. Pretty sure the lock won't last in this configuration. Attached plot for ASC DHARD pitch shows all of the increased signal is in the .44 hz quad mode. Red is from 4 hours ago, blue is current.

Images attached to this comment
jim.warner@LIGO.ORG - 21:27, Thursday 06 April 2017 (35375)ISC

We just lost lock. I think we should re-engage the oplevs.

Images attached to this comment
brett.shapiro@LIGO.ORG - 22:33, Thursday 06 April 2017 (35376)

Recall that top to top transfer functions to do not show the damping of the test mass. They can mislead you that there is high damping when there isn’t really at the test mass. So all transfer functions in this case should show the test mass as an output. Input probably isn’t too important, but from a consistency point of view may as well make all transfer functions between the same stages.

I agree that doing the oplev damping in hard and soft coordinates may be better than local coordinates. In local coordinates, all test masses need high bandwidth damping to accommodate the hard modes. In hard and soft coordinates, only the hard loops need high bandwidth damping. So that might give us a win of up to sqrt(2) in noise for the same amount of damping.

jim.warner@LIGO.ORG - 22:52, Thursday 06 April 2017 (35377)ISC

After getting relocked, the range started dropping again and the ASC foms showed .44hz mode was still ringing up even with oplev damping. After talking to Sheila, we realized the CSOFT gain wasn't in the guardian yet (it was still at .2, should be .4 with oplev damping off), so I set 30 second tramps on the CSOFT P and oplev damping and with some quick clicking ramped the CSOFT gain up and the oplev gains back to zero.  The lock survived this, ASC foms settled down and the range started recovering. I've accepted the changes in SDF. Hopefully the lock will survive the night, but if it doesn't the CSOFT gain will show up in SDF. It should be set to .4.