The h1isiham3 model running on h1seih23 was quite old, from June 8th. Before we start ISI testing we (Hugo and I) upgraded this model by bringing in the latest L1 HAM3 model and converting L1->H1. From now onwards we should maintain the h1isiham3 model as separate from the l1isiham3. The new model required new Dolphin IPC channels from the h1susmc2 model, so that model was rebuilt from L1 but was not restarted on h1sush34. I had some problems recompiling the h2suspr2 model, there are IPC errors between MC2 and PR2 which are currently unresolved.
There are still some errors with the H1 ISI HAM3 medm screens. There appears to be residual links to M1 and S1 systems. I'll work with Hugo to resolve the H1 screens.