3:41 Lockloss

3:41:  Lockloss

PSL Issues:  (I marked some CORRECTIVE MAINTENANCE time, but not sure this is FRS-worthy):

Corresponding with this lockloss, had the following PSL Guardian Messages:

For ISS Diff Power (it was oscillating between 3-14), I turned OFF the Autolocker, clicked Manual & I thought I turned it OFF---so I could tweak the Ref Signal, but I didn't have luck here.  I tried restoring everything.

Since Noise Eater was RED/Off, I went out to the LVEA to toggle the switch.  This made the Guardian Messages go away, and I went about locking.

SDF Diffs & Servo Gain Changes

Upon reaching NLN, had a few diffs (attached).

PSL Power Drifting Down, Servo Gains, & ACCEPTING Diffs Until Alignment Can Be Tweaked Up

Ended up calling Keita about these Servo Gains before attempting to go to OBSERVING.

Looks like the PSL power has been dropping over the last month (see attached IMC trends) and we finally took a jump down to where the REFL servo gains were increased & noticed by SDF (see attached trends of ISS diff power, PSL Power, & IMC/LSC REFL SERVO IN1GAINS).  

We also looked at the DARM (H1:CAL-DELTAL_EXTERNAL_DQ) spectra & compared tonight (blue) to March 19 (red).  Since the high freq isn't getting increasingly noisy with higher frequency, we are OK with the SERVO gain values.  

So, if we have more of these DIFFs over the weekend, we should ACCEPT them in SDF.  Hopefully we can last until Peter/Jason can touch up the alignment for us next week.

5:20 Back to OBSERVING