nutsinee.kijbunchoo@LIGO.ORG - posted 16:24, Thursday 20 April 2017 (35696)
4.7Hz violin mode damping turned off

The 4.7kHz line was reasonably high, but wasn't ringing up. Evan and Sheila thought it might be corrupting the data so we tried to damp it. The original setting of +30 deg phase didn't do much so I turned on -60deg filter as Patrick suggested. The line went down on DARM ASD but rang back up very quickly. I didn't have my dtt opened but I believe what happened was -60 + 30 deg phase worked well with one of the lines but quickly rang up the other line. I turned off the -60deg and was able to damped the other line with +30 deg. Right now the gain is left at 0. I accepted the SDF diff and went back to Observe.

 

If 4.7kHz doesn't ring up by the end of EVE shift, I would like to comment out the line in the guardian that turn on the gain for this mode.