Displaying report 1-1 of 1.
Reports until 09:36, Monday 24 April 2017
H1 PSL
edmond.merilh@LIGO.ORG - posted 09:36, Monday 24 April 2017 (35748)
PSL Weekly Status Report - FAMIS 7435

Laser Status:
SysStat is good
Front End Power is 34.07W (should be around 30 W)
HPO Output Power is 168.5W
Front End Watch is GREEN
HPO Watch is GREEN

PMC:
It has been locked 2 days, 7 hr 33 minutes (should be days/weeks)
Reflected power = 16.55Watts
Transmitted power = 64.53Watts
PowerSum = 81.08Watts.

FSS:
It has been locked for 0 days 10 hr and 24 min (should be days/weeks)
TPD[V] = 3.623V (min 0.9V)

ISS:
The diffracted power is around 2.8% (should be 3-5%)
Last saturation event was 0 days 10 hours and 29 minutes ago (should be days/weeks)

Possible Issues:
none

Displaying report 1-1 of 1.