H1 SUS (CAL, CDS, DetChar, ISC, OpsInfo, SYS, VE)
jeffrey.kissel@LIGO.ORG - posted 13:54, Tuesday 25 April 2017 (35778)
Charge Measurement Update; New Plan -- Same as the Old Plan
J. Kissel

I've taken this week's effective bias voltage measurements that track accumulated charge on the test mass/reaction mass electrostatic drive system. 

The standard metrics:
- Monitoring the single-frequency actuation strength of each quadrant as a function of requested bias voltage using Pitch and Yaw optical optical lever signals to measure the response. 
- Monitoring the overall longitudinal strength comparing an ESD excitation against a photon calibrator excitation at adjacent single frequencies.

The conclusions from both these metrics have not changed from the last two measurements (LHO aLOGs 35553, 35366):

Just before we vent on May 8th, during last minute preparations as we bring the IFO down, let's 
    - Turn OFF the ETMX ESD bias completely, and leave it OFF 
    - Leave the ETMY bias ON at +400 [V], with the opposite sign as in observation (Negative in NLN, switch to Positive for vent)
for the duration of the vent (i.e. until we open back up to the arms)
Images attached to this report