Displaying report 1-1 of 1.
Reports until 15:41, Thursday 27 April 2017
H1 AOS (AOS, SEI, SUS)
jeffrey.bartlett@LIGO.ORG - posted 15:41, Thursday 27 April 2017 - last comment - 16:04, Thursday 27 April 2017(35845)
OpLev 7-Day Trends
   Post the 7 day OpLev trends. 

SUM: ITMY - No data after 04/24 - Zoomed the plot way out and could find a bottom.
     ETMY - Big jump after the 04/25 maintenance window. This is a result of Jason's tune up during the window. 

Yaw: ITMX - Near the 10urad line. Trend line is flat.

Pitch: SR3 - Near the 10urad line. Trend line is flat.   
Images attached to this report
Comments related to this report
jason.oberling@LIGO.ORG - 16:04, Thursday 27 April 2017 (35847)

Everything looks normal with the trends.  ITMx yaw is beginning to get close to the 10 µrad limit, and will likely need re-centering in the near future.

Regarding ITMy SUM, it changed due to my swap of the laser (reported here).  Unfortunately it's not entirely out of the woods yet, as the laser is showing signs of low output power.  While the SUM signal doesn't show up on Jeff's plot above (not sure if this is due to something in the script or some odd artifact of the OS change with the control room workstations, or something else entirely), it is there; DetChar summary page for ITMy oplev shows it, as well as the oplev MEDM screen.  The oplev seems to be functioning normally, just looks like its output power is low.  I plan on investigating this during the 5/2/2017 maintenance window.

Displaying report 1-1 of 1.