LHO VE
chandra.romel@LIGO.ORG - posted 19:50, Tuesday 09 May 2017 - last comment - 22:07, Tuesday 09 May 2017(36114)
Vacuum summary (vertex vent)
Comments related to this report
kyle.ryan@LIGO.ORG - 22:07, Tuesday 09 May 2017 (36115)
I double checked the mechanical pressure gauges downstream of the 1.5 psi pressure regulator this morning after we noticed a high purge pressure value for the Vertex.  The flow demand was high at this time as BSC3's door was off (HAM4 door too?) and these gauges were showing about 1/2 psig which is as expected for this flow demand.  We resolved to adjust the "ATM" trim pot on PT120 before pumping tomorrow.  

I now am wondering if the pointers of these low pressure mechanical gauges were rotated past 360 degrees - indicating 1/2 psig but actually being 1/2 psig past n(360)?  We need to close off the purge air supply and allow the Vertex to "blow down" and equalize to room pressure tomorrow to see what PT120A reads in orderto see what is going on.  