Replacement of ITMy In-Vacuum Hartmann Lens on WHAM4 Optics Table

(Aidan B, Corey G, TJ S, Jim W, PLUS much assistance to by Vacuum Team, Betsy W for prep work, & Jeff B. for staging of Contamination Control supplies)  

All times in PDT

On Tues (May9th) multiple in-vacuum tasks were on the docket for work on H1, and this log covers the work at/inside HAM4:  Replacement of the H1 HWSY VAC LENS.  While performing this work, we followed Vent Plan (E1700124) & followed STEPS #32 - 39.  Also used the Chamber Entry & Exit Procedure (E1201035, sheet4).  Attached are some photo highlights from this job (all photos are on resourcespace [links below]).  

HAM4 was roughly open from 10am - 2pm with only minimal entry within the chamber (access was from the North & only arms/upper body entered the chamber for this work task).

NOTE:  During this activity there was an Alert/Site Area Emergency event on the Hanford site (first time this has happened for LHO in ~20years of being here).

Summary of HAM4 Activities:

  0.3um (particle/L) 0.5um (particle/L)
Before Door Removed 68 12
After Door Removed 5 1
In-Chamber 11 4
Images attached to this report
Comments related to this report
aidan.brooks@LIGO.ORG - 08:23, Thursday 11 May 2017 (36132)

The new lens is from the bag (and box) labeled: CVI PLCX-50.8-360.6-UV-700-900, per Table 2 of T1000179-v20

Images attached to this comment
aidan.brooks@LIGO.ORG - 08:36, Thursday 11 May 2017 (36133)

The old lens had a label penciled onto the side. This read: PLCC-50.8-772.6-UV

This confirms our educated guess back in October 2014: (T1400686, page 2).

 

Images attached to this comment