Reports until 12:10, Friday 12 May 2017
LHO VE
logbook/robot/script0.cds.ligo-wa.caltech.edu@LIGO.ORG - posted 12:10, Friday 12 May 2017 - last comment - 14:06, Friday 12 May 2017(36161)
CP3, CP4 Autofill 2017_05_12
Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 473 seconds. TC B did not register fill. LLCV set back to 20.0% open.
Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 208 seconds. TC A did not register fill. LLCV set back to 45.0% open.
Images attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 14:06, Friday 12 May 2017 (36165)

Raised CP3 to 21% open.