LHO VE
chandra.romel@LIGO.ORG - posted 11:45, Friday 12 May 2017 - last comment - 16:14, Friday 12 May 2017(36159)
Vertex vacuum pressure and other updates

 

 

WARNING:  don't turn on high voltage until pressure reads e-6 Torr range on PT-120B gauge

Images attached to this report
Comments related to this report
john.worden@LIGO.ORG - 12:11, Friday 12 May 2017 (36160)

A reminder - Paschen's curve of breakdown voltage vs pressure. And note that we have gaps at feedthroughs in the mm range rather than cm. 

From Wikipedia;

https://en.wikipedia.org/wiki/Paschen%27s_law

 

Images attached to this comment
chandra.romel@LIGO.ORG - 16:14, Friday 12 May 2017 (36171)

GV4 annulus vacuum is surviving on the IP alone at 5 mA. I left the turbo hardware connected, but everything is powered OFF.

Valved out water from QDP80 pump and turned booster pump off. Left x-beam manifold turbo station energized until I leave site.

Vertex turbo inlet pressure is 2.1e-6 Torr. BSC2 pressure is 7.9e-6 Torr.