CP3, CP4 Autofill 2017_05_19
Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 274 seconds. TC B did not register fill. LLCV set back to 21.0% open.
Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill not completed after 3600 seconds. LLCV set back to 33.0% open.
Images attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 14:47, Friday 19 May 2017 (36291)

Manually filled CP4 from control room at 100% open. Took 11 minutes. Lowered to 36% open (from 33%).