Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 274 seconds. TC B did not register fill. LLCV set back to 21.0% open. Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill not completed after 3600 seconds. LLCV set back to 33.0% open.
Manually filled CP4 from control room at 100% open. Took 11 minutes. Lowered to 36% open (from 33%).