Kyle 
(Tasks part of WP #6644)
~1015 hrs local -> I valved-in the Vertex RGA to the combined vacuum volumes of the Vertex+YBM+XBM while these volumes were being pumped by their respective turbo pumps.  PT120's response is attached (1 hour of second data).  "From my head" 10-9 torr/10-5 L = 10-4 torr/L of gas accumulated in the unpumped ~5L RGA volume in 14 days?  So, no significant air leaks in the RGA joints.
~1045 hrs. local -> I energized the Vertex RGA filament (note: I am confused with the Pfeiffer software's naming of my connect/scan session with the included phrase "Airdemo..".  The other versions of this software don't do this and I makes me question as to whether I am actually connected to the Vertex RGA (h0rgacs) or am I connected to their "Simulator" or virtual RGA in this connect/scan session?