I performed calibrations of both end stations this week as a target of opportunity during the pumpdown of the CS. I forgot to turn off the CW injections for End X, so the data we got is not good for calibration. I will revisit End X at a later date. End Y was successful though and the full report can be found here. In summary, the RxPD force coeffiecient is within 0.24% and the TxPD force coefficient is within 0.22% of the values used for O2 calibration.