Displaying report 1-1 of 1.
Reports until 21:55, Friday 26 May 2017
H1 General
patrick.thomas@LIGO.ORG - posted 21:55, Friday 26 May 2017 - last comment - 16:31, Saturday 27 May 2017(36464)
Sitting at 26 Watts, stable so far
Sheila D., Patrick T.

Reached 26 W at approximately 04:13 UTC. Increased CSOFT gain by 3dB. This seems to have improved the stability. Lines just above 20 Hz are from a2l script. Frequency noise is high above 2kHz. Jitter noise appears better.
Images attached to this report
Non-image files attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 22:35, Friday 26 May 2017 (36465)
  • After resetting the alignment references earlier today, we had several lockloses in the state CARM 5PM. Patrick and I got through it once by going slowly, so I have increased the ramp time from 5 seconds to 15.
  • The alignment references do seem good: we locked with a recycling gain of 32 and the ASC came on without any problems or pulling the alignment around much. 
  • We increased the CSOFT gain because our first 26Watt lock tonight we lost because of the CSOFT instability.  This is in the guardian, the measurement attached shows the reference (gain 0.4 and the new measurement gain 0.6)
  • The jitter coupling is much better than it was before the vent.  You can compare the coherence that Patrick posted above to the one that Keita posted 
  • The frequency noise is much worse.  
  • We ran a2l and after we were satisfied that we weren't going to see the CSOFT instability again we moved on to low noise. We lost lock right after the beam diverters closed, although it could have been due to an alignment problem. 

The second screenshot attached is the same as Patricks plot with the best spectrum that we got just before loosing lock. 

 

Images attached to this comment
Non-image files attached to this comment
aidan.brooks@LIGO.ORG - 09:17, Saturday 27 May 2017 (36467)TCS

Here are the Hartmann sensor measurements from this time. The ITMY thermal lens looks to be around 10cm in diameter, or roughly the same spatial scale as the IFO beam. This, and the lack of any strong localized features, is consistent with only uniform absorption on this optic. 

ITMX shows the point absorber again.

Note: we made every effort to center the HWS beams on the optics. As yet, I have not quantified the exact center of each optic in the HWS coordinate system. The plotting code defaults to assuming that the HWS beam is exactly centered but it is feasible that we may be off center by a couple of centimeters.

ITMY-HWS after around 10 minutes

The contour lines are spaced 8nm apart.

 

ITMY after 30 minutes

The double-pass thermal lens here is around 28 uD. The single-pass thermal lens is about 14 uD.

 

ITMX after ~5 minutes

Images attached to this comment
kiwamu.izumi@LIGO.ORG - 16:26, Saturday 27 May 2017 (36470)

Looking at the Seidel coefficients of HWSs X and Y, I see that our TCS pre-loading scheme on ITMY is not doing a great job. It leaves a spherical power of 15 uD as we remove the pre-loaded CO2 power. See the attached. I am not sure why HWSX data is noisier than HWSY.

Images attached to this comment
aidan.brooks@LIGO.ORG - 16:31, Saturday 27 May 2017 (36471)

I need to implement a weighted best fit parameter estimator for the HWS Python code. Chances are that HWSX has some noisy centroids that are skewing the measurement. When we weight the parameter fitting by the inverse of the variance of each centroid, we should get a much cleaner estimation of the spherical power.

I will implement that this week.

Displaying report 1-1 of 1.