Issue with watchdog tripped - Richard, Filiberto, Hugo
Corrected, can now actuate
We managed un-tripping the coil driver chassis bio watchdog of HAM3 by:
In Vacuum cabling - Corey, Hugo
CPS H3 and CPS V3 were swapped - corrected
I took spectra on our sensors this morning and noticed H3 and V3 were swapped. Corey went looking at it and located it on the in-vaccum side of the feedthrough.
The binary word that should be used for HAM3, in ISI_CUST_CHAMBER_BIO.adl is:
65535