CP3 and CP4 auto overfills are now programmed to run daily instead of thrice-weekly (Mon, Wed, Fri).
CP3:
Chandra requested that today's CP3 auto overfill not run. This has been taking out of the crontab until further notice.
CP4:
CP4 ran at 11:02 PDT with the new configuration: daily run, cut-off temperature lowered from -60C to -100C. The attached trend shows CP4 thermocouple second trend data for 20 minutes for the last three runs (Mon 5/29, Wed 5/31 and Thu 6/1).