H1 SEI
hugo.paris@LIGO.ORG - posted 18:28, Monday 30 July 2012 - last comment - 11:55, Tuesday 31 July 2012(3659)
HAM3-ISI - In Chamber Testing Progress - Transfer Functions Running Overnight

Filiberto, Richard, Corey, Hugh, Hugo,

 

Today's effort brought us to the point where we are able to run transfer functions:

Transfer Functions are running overnight. They should be done by 6am.

 

Note: There is a mailing list to receive the alerts regarding these transfer functions. Please let me know if you would like your email address to be added to it.

Comments related to this report
hugo.paris@LIGO.ORG - 09:52, Tuesday 31 July 2012 (3662)

The computer overnight TFs were running on froze (black screen) around 11.30pm. I was able to retrieve the TF sections measured until then.

TFs were compared to the ones measured on LLO HAM3. Accordance regarding result on CPSs is good (plots attached).

Results on GS13s appeared filtered differently than LLO and need to be analyzed in more details.

GS13s were tested. Their spectra taken with when the ISI is tilted (most critical configuration for GS13s) show that they work properly (see attached plots).

Images attached to this comment
hugo.paris@LIGO.ORG - 11:55, Tuesday 31 July 2012 (3664)

LLO-HAM3 Transfer Functions used for the comparison plots (CPS-H, CPS-V) were recorded in-chamber, in-air, on January 18th 2012. These transfer functions are the one presented in LLO-HAM3 phase II testing report under the section 2.2.1 of the initial chamber testing chapter.