kiwamu.izumi@LIGO.ORG - posted 17:21, Friday 02 June 2017 - last comment - 18:15, Friday 02 June 2017(36629)
Update on ALS Y issue

Fil, Daniel, Jenne, Keita, Patrick, Richard, Sheila, Kiwamu,

The investigation continues. No solid conclusion or resolution yet.

Comments related to this report
sheila.dwyer@LIGO.ORG - 18:15, Friday 02 June 2017 (36631)

Just to summarize what we think we know so far:

Things that were exonerated by bypassing PLL test:

  • PDH sensing (diode, demodulator (until it broke), LO, cables)
  • Laser (temperature wasn't very different between PLL bypass and normal locking configuration lock)
  • PLL servo board and beckhoff control cables.

Other things that we think are exonerated:

  • PDH servo board, because we swapped the board and still have glitches
  • loop instabilities, I measured both loops this morning in the normal configuration.  PLL has a bandwidth of 27kHz and 43 degrees of phase, PDH has 2.5kHz ugf and 66 degrees of phase
  • The PSL laser shouldn't be the problem, because if it were we would see the problem at End X.
  • The beat note from the fiber BBPD looks like a nice clean sine wave

Things that we haven't eliminated:

  • PFD- we could try replacing this with a mixer as a test
  • VCO- we could try to replace this with an IFR as a test
  • BBPD
  • PDH servo board beckhoff cables- we could try to replace these with a CM board tester.  Keita calling to try to find out where one of these is stored.