(not Bram)
Attached is a plot of the response of the laser crystal TEC when stepping 13mK up and then down at the slow input (assume -1K/V sensitivity). The response shows a spike by about +9mK and then -9mK (not clear how good either calibration is). The response time is roughly 5 sec which would indicate a ~30 mHz pole in the slow input of the laser.