Displaying report 1-1 of 1.
Reports until 16:02, Tuesday 20 June 2017
H1 CDS (CDS)
sheila.dwyer@LIGO.ORG - posted 16:02, Tuesday 20 June 2017 - last comment - 16:52, Tuesday 20 June 2017(37037)
DTT abort lockloss?

I think that this lockloss was caused by DTT suddenly outputting a large excitation when I don't think it should have, at 1182033600.  I increased the amplitude of the noise excitation in the attached template from 10000 to 20000, the template has a 5 second ramp up time, but for some reason the excitation reached an amplitude of 45,000 counts very suddenly as shown in the attached screenshots. 

Since the input to the filter was off, the only thing coming into this filter should have been the excitation.  

Images attached to this report
Non-image files attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 16:52, Tuesday 20 June 2017 (37041)

This happened again, this time I just let the measurement finish and when it was ramping off the interferometer unlocked. Screenshots of the two incidents are attached.

 

Images attached to this comment
Displaying report 1-1 of 1.