Reports until 20:47, Friday 23 June 2017
H1 General (CAL)
thomas.shaffer@LIGO.ORG - posted 20:47, Friday 23 June 2017 (37103)
Ops Mid Shift Report

Locked for 34 hours at 67Mpc.

DIAG_MAIN is showing "PCAL: Y RX PD is greater than 1% off". Trending the channel that is looks at (H1:CAL-PCALY_RX_PD_VOLTS_OUTPUT), it has dropped very slightly. This same message showed up about 12 hours ago, so I got suspicious of temperature. Sure enough, you can clearly see it dip with the temperature. See 2 day trend below

Images attached to this report