H1 SEI
hugh.radkins@LIGO.ORG - posted 13:28, Wednesday 25 July 2012 - last comment - 10:10, Friday 03 August 2012(3586)
H1 HAM3 HEPI Actutors Attached & Released--IAS says OK! Prt Duex
OK now we have all the HEPI Actuators attached & released at the correct elevation and IAS has signed off on the position.

Elevation & level is 0.1mm below nominal of -250.5mm (wrt LIGO Global) with a +-0.1mm levelness.

Attached is my field notes of the final dial indicator readings and the level survey-enjoy.

Thanks to Mitchell & Jason
Non-image files attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 10:10, Friday 03 August 2012 (3712)
I misspeak above where I say wrt LIGO Global.  The number above, -250.5mm is the LIGO Global elevation corrected to the local leveling field.  The HAM ISI Optical Table is positioned to -252.9mm.  2.4mm is the correction to local level hence the -250.5mm; please pardon the confusion. -H