Moved the roaming STS2 to position 7 and here report response during wind of 3 July. Not as windy as usual studies, only about 15 mph rather than 20. The wind was at 45degrees, that is directly from true South, or, from +Y-X.
Similar to Roam6, the response at Roam7 indicates this position is quieter than the ITMY location, especially in the X dof. Since this position move was parallel to the 'south' wall wrt Roam6, it makes some sense that the X dof response is nearly identical to the response to wind at positions 6 and 7. Maybe time to jog north (away from south wall.) Also, moving away from the 'pinning' potential of BSC4 does nothing to suggests a real pinning effect.