PSL Weekly Report (FAMIS #7447)

Laser Status:
SysStat is good
Front End Power is 33.96W (should be around 30 W)
HPO Output Power is 154.9W
Front End Watch is GREEN
HPO Watch is GREEN

PMC:
It has been locked 0 days, 7 hr 19 minutes (should be days/weeks)
Reflected power = 16.12Watts
Transmitted power = 57.58Watts
PowerSum = 73.7Watts.

FSS:
It has been locked for 0 days 1 hr and 44 min (should be days/weeks)
TPD[V] = 2.82V (min 0.9V)

ISS:
The diffracted power is around 3.2% (should be 3-5%)
Last saturation event was 0 days 1 hours and 44 minutes ago (should be days/weeks)

Possible Issues:

No issues to report