This is a quick summary of work done by many people today:
The last plot shows a couple of sensitivity curves, illustrating the problem.
Josh, TJ, Andy, Beverly,
A status update on Detchar's efforts to follow up the excess noise reported above. We haven't found anything clear yet, but there are some small oddities.
The spectra posted above look strange because the DARM noise below 30Hz is lower now than before while above 30Hz it's higher - a hint perhaps. Perhaps a filter or switch not engaged?
We ran Bruco: https://ldas-jobs.ligo.caltech.edu/~thomas.massinger/bruco_H1_July19_10UTC/
No smoking gun broadband coherence around 30-80Hz. Notable coherent things are:
We checked that h(t) and DARM IN1 both see this noise (so not a calibration filter mismatch) (yellow and blue are now/bad times).
We started looking for a switch that says it's switched but isn't like in LLO's 33071, but there one FASTIMONs was way higher than the others, and so far we don't see that. We checked all of the L2/M2/M3 FASTIMONs and nearly all of them are the same now as they were for two quiet reference times with good range (30th of June and 4th of July). Some exceptions are (yellow and blue are now/bad times)
We checked all SUS BIO MON channels and they are all in same state as the reference good times.