Earthquake, Lockloss, Seismon stopped
M 6.5 - 36km W of Yongle, China
2017-08-08 13:19:49 UTC 33.217°N   103.843°E 10.0 km depth

Lockloss at 14:05 UTC.

It seems the System Time for Seismon stopped updating at 1186191618 GPS, Aug 08 2017 01:40:00 UTC, yet its System Uptime and Keep Alive are still updating.

14:10 UTC ISI platforms started tripping. Transitioned ISI config to LARGE_EQ_NOBRSXY.
Tripped:
ISI ITMY stage 2
ISI ETMY stage 2
ISI ITMY stage 1
ISI ITMX stage 1
ISI ITMX stage 2
ISI ETMY stage 1
SUS TMSY

Leaving ISC_LOCK in DOWN. Starting maintenance early.

14:18 UTC Set observing mode to preventive maintenance.
Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 08:42, Tuesday 08 August 2017 (38070)

I don't quite know what happened with seismon. I updated the USGS client yesterday and restarted that code, but I didn't make any changes to the seismon code. I've restarted the seismon_run_info code that we are using and that seems to have fixed it. Maybe the seismon code crashed when I added geopy yesterday?