Reports until 12:28, Tuesday 08 August 2017
H1 General
cheryl.vorvick@LIGO.ORG - posted 12:28, Tuesday 08 August 2017 - last comment - 16:08, Tuesday 08 August 2017(38081)
Maintenance:

Maintenance tasks: all times UTC

Current:

Comments related to this report
cheryl.vorvick@LIGO.ORG - 12:38, Tuesday 08 August 2017 (38083)
  • 19:30 - Kyle done unloading at CS
  • 19:31 - Travis to EX to gather tools
  • 19:34 - JimW fixing fiber polarization
jeffrey.kissel@LIGO.ORG - 16:08, Tuesday 08 August 2017 (38089)
19:46 Fil to CER to pick up his notebook.
19:51 Fil out
19:50 Beginning Initial alignment.
19:55 Shiela and Thomas to End stations to revert ETMX ESD Cabling
20:04 Jason to Mid X
20:30 Jason back 
20:45 Having trouble keeping green arms locked, once ALS WFS + CAM are engaged, even with good flashes to start
      Found Green ITM camera image processing had stalled since h1script0 was rebooted.  
21:00 Fire Alarm
21:15 Guardian node errors as a result of failed reverting of attempt to lock ALS on ETMY in the morning

Sheila and Thomas also went to EX to confirm that ESD EX is cabled up correctly.
While snooping, they found that C and D fault lights were on for the TMSX OSEM sat amps. Thinking this was a problem, we restarted the coil drivers twice (no change), then unplugged then reseated the CD to SatAmp cables at the satamp (no change). Then we remembered that the TMS's second top mass signal chain only runs 2 OSEMs on channels A and B, so C and D will always have fault lights on.
All cables and power restored, TMSX returned to normal functionality.