J. Kissel Was able to grab an entire suite of calibration measurements this afternoon. Data files listed below, will process in due time. Preliminary results are as expected -- attached is a screen shot of "perfect" reference calibration against current CAL-DELTAL_EXTERNAL, which we know is *not* corrected for time dependent correction factors -- namely kappa_TST which is up at ~10% at the moment, likely causing the shown discrepancy. Actuation: /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Measurements/FullIFOActuatorTFs/2017-08-16 2017-08-16_H1SUSETMY_L1_iEXC2DARM_25min.xml 2017-08-16_H1SUSETMY_L1_PCAL2DARM_8min.xml 2017-08-16_H1SUSETMY_L2_iEXC2DARM_17min.xml 2017-08-16_H1SUSETMY_L2_PCAL2DARM_8min.xml 2017-08-16_H1SUSETMY_L3_iEXC2DARM_8min.xml 2017-08-16_H1SUSETMY_L3_PCAL2DARM_8min.xml Sensing: /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Measurements/SensingFunctionTFs 2017-08-16_H1DARM_OLGTF_4to1200Hz_25min.xml 2017-08-16_H1_OMCDCPDSUM_to_DARMIN1.xml 2017-08-16_H1_PCAL2DARMTF_4to1200Hz_8min.xml 2017-08-16_H1_PCAL2DARMTF_BB_5to1000Hz_0p25BW_250avgs_5min.xml