Reports until 15:03, Monday 28 August 2017
H1 General (SUS)
edmond.merilh@LIGO.ORG - posted 15:03, Monday 28 August 2017 (38405)
Balance Calibrating Top Gun Deionizing N2 Guns

The following units have been balance calibrated as per LIGO-T1600116-v1:

#132740069233007, #132740069233002 and 132140069233006

The unit ending in 007 had been currently in use on the optic in the LSB Bonding lab and was found to be out of spec as defined in the procedure listed above. I was able to adjust it into an acceptable balance as per procedure.

#132740069233004 failed to meet the tolerances as stated in the above procedure.

There is another unit in the Bonding Lab garbing area that has no regulator. I didn't get the number off that one

The data for these units may be found here