jason.oberling@LIGO.ORG - posted 10:03, Tuesday 29 August 2017 (38413)
PSL Weekly FAMIS Tasks (FAMIS 3665 & 8437)

This morning I completed the weekly PSL FAMIS tasks.

HPO Diode Current Adjustment (FAMIS 8437)

With the ISS turned OFF, I adjusted the operating current of the HPO pump diode boxes.  DB1 increased by 0.2A and no change on DB2, DB3, and DB4.  The changes are summarized in the below table; I also attached a screenshot of the PSL Beckhoff main screen for future reference.

  Operating Current (A)
Old New
DB1 49.8 50.0
DB2 52.7 52.7
DB3 52.7 52.7
DB4 52.7 52.7

I also adjusted the operating temperatures of the DBs; changes summarized in the below table:

  DB1 DB2 DB3 DB4
Old New Old New Old New Old New
D1 28.5 28.5 20.0 19.5 21.0 21.0 23.5 23.0
D2 28.5 28.5 19.5 19.0 25.0 25.0 21.0 20.5
D3 28.5 28.5 20.5 20.0 25.0 25.0 22.5 22.0
D4 28.5 28.5 18.5 18.0 22.0 22.0 21.0 20.5
D5 28.5 28.5 18.5 18.0 26.0 26.0 23.0 22.5
D6 28.5 28.5 19.0 18.5 20.5 20.5 23.0 22.5
D7 28.5 28.5 19.5 19.0 21.5 21.5 23.0 22.5

The HPO is now outputting 154.9W and the ISS is back ON.  This completes FAMIS 8437.

PSL Power Watchdog Reset (FAMIS 3665)

I reset both PSL power watchdogs at 16:43 UTC (9:43 PDT).  This completes FAMIS 3665.