Second Cycle of ETMX TMDS Complete
J. Kissel, N. Mavalvala, K. Ryan, D. Sigg

We've successfully completed the second cycle of the Test Mass Discharge System on H1 SUS ETMY (BSC9/BSC5). 

Operation went much more smoothly today that yesterday. We were open to the chamber sending ions in by 08:00a PDT, and valved out by 08:30a PDT. We decided only to go up to a BSC chamber pressure of 15 [Torr], partially due to person-power constraints, but mostly because the bulk to the ion neutralization happens when the TMDS is first valved in; as more clean air is in the chamber, the ionic mean free path to the test mass decreases. This should also help us a little on pump-down time. All other settings were identical to what worked well yesterday (see LHO aLOG 38430):
   - clean air flow rate of ~70 [L/min] (or "slm")
   - ion "current" (voltage) occasionally measured by the electrometer with high-amplitude square wave (12Vp amplitude) of +/- 8 [V]
   - pressures inside the TMDS of ~100 [Torr] as measured by the baratron
Attached are the raw notes (but they're not very exciting -- just 4 measurements of the ion "current" during the 1/2 hour discharge cycle).

Based on the current pump-down rate, we expect to perform our third and final cycle around 01:00p PDT today. 
Non-image files attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 11:19, Wednesday 30 August 2017 (38454)VE
BSC5/BSC9 Pressure trend to back up my prediction of 01:00p PDT third cycle readiness.
Images attached to this comment