Reports until 17:52, Wednesday 30 August 2017
LHO VE
kyle.ryan@LIGO.ORG - posted 17:52, Wednesday 30 August 2017 - last comment - 22:30, Wednesday 30 August 2017(38462)
Pumping X-end with Turbo overnight
Kyle, John 

Two additional discharge cycles, in addition to yesterday's single cycle, were completed by D. Sigg, J. Kissel and N. Mavalvala today.  Following the final gas admission, and with the X-end pressure at ~16 Torr or so and the TMDS port isolated, John and I decoupled the Surface Disharge Ionizer from BSC9 and installed a 2.75" blank in its place.  I then opened the 1.5" TMDS port gate valve and "burped" these few ccs of room air into the X-end vacuum volume.  This valve was then left open.  

I roughed down using the QDP80 to 0.5 Torr < pressure < 1.0 Torr before spinning up the Turbo.  This is the same approach as was used for the previous two pump downs.  

I am leaving the X-end on the Turbo overnight and plan to execute WP #7134 in the morning.  
Comments related to this report
kyle.ryan@LIGO.ORG - 20:39, Wednesday 30 August 2017 (38463)
Factor of 10 between gauges.  PT510 is a 20 yr. old cold cathode and PT525 is 2 yr. old Bayard-Alpert parani wide-range.  I'll think about this tomorrow ;)
kyle.ryan@LIGO.ORG - 22:30, Wednesday 30 August 2017 (38464)
It looks like PT525 is non-changing?