H1 PSL
jason.oberling@LIGO.ORG - posted 09:53, Tuesday 05 September 2017 (38513)
PSL Weekly FAMIS Tasks (FAMIS 3666 & 8438)

Today I completed the weekly PSL FAMIS tasks.

HPO Pump Diode Current Adjust (FAMIS 8438)

With the ISS turned OFF, I adjusted the operating current of the HPO pump DBs, changes are summarized in the table below.  I also attached a screenshot of the PSL Beckhoff main screen for future reference.

  Operating Current (A)
Old New
DB1 50.0 50.2
DB2 52.7 52.8
DB3 52.7 52.8
DB4 52.7 52.8

I also adjusted the operating temperature of DB1: all diodes were running at 28.5 °C, they are now operating at 28.0 °C.  This ISS is now back ON and the HPO is outputting 154.9 W.  This completes FAMIS 8438.

PSL Power Watchdog Reset (FAMIS 3666)

I reset both PSL power watchdogs at 16:42 UTC (9:42 PDT).  This completes FAMIS 3666.