borja.sorazu@LIGO.ORG - posted 10:53, Thursday 14 September 2017 (38633)
Ringdown Q measurements of the 6th and 7th harmonic for the QUAD violin modes - After Earthquake

These ringdown measurements, of the LHO QUAD suspensions violin mode 6th and 7th harmonics, were obtained 2 days after the big Earthquake hitting LHO on the 20170706. For this analysis I used 17 hours of detector data in Low noise state from "20170708 06:30:00" UTC. None of the 6th and 7th harmonics were being actively damped or excited. Although clearly the Earthquake excited these harmonics well enough to get very nice ringdowns.

In this analysis, a line tracker (iWave) was applied over each of the identified 6th and 7th harmonic frequencies, locking onto them.

I give next the results for those frequencies that show ringdowns. I also attach 'png' plots which shows; in each column the mode monitored with the top plot being the frequency tracked as a function of time, the middle plot is the 'log(Amplitude)' (natural logarithm of the mode's amplitude). The red dashed lines are respectively the median of the tracked frequency and the fitted first order polynomial to the 'log' of the mode's ringdown. The plot at the bottom shows the Phase deviation respect to the linear fit.

6th Harmonic

Mode frequency (Hz)          Q  

2836.347                           792664427

2837.676                           735427842

2839.189                           667931233

2841.245                           645741246

2842.507                           503371109

2843.141                           669024332

2843.431                           659048452

2844.486                           715255321

2846.824                           675302066

2847.020                           608809212

2848.206                           659351953

2849.444                           534289274

2849.583                           405660077

2850.597                           574906401

2853.482                           712952420

2853.710                           401649469

2857.727                           724216344

2858.379                           624072692

2864.457                           770311177

2866.174                           671254082

2868.132                           427751583

2878.644                           498367912

2878.701                           429105364

2880.567                           621008999

2880.854                           614386214

2882.491                           585658907

 

The 7th harmonic values will be added below in the comments.