H1 ISI CPS Sensor Noise Spectra Check - Weekly FAMIS #6915

Everything looks normal save for HAM3 which shows a slight elevated CPSINF_H2.

Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 14:04, Thursday 14 September 2017 (38636)

Thanks Ed.  Yes even on the thumbnails it is pretty clear that the H2 sensor on that plot is not like the others.

I unseated and reseated the H2 card in the satellite rack and the noise returned to normal; but, only for an hour or so.  Next I'll try the interface chassis power cycle.