It looks like the only "hiccuup" experienced by the vacuum system following the site wide power outage was that IP6's controller defaulted to a "standby" state, i.e. High Voltage was disabled -> I re-enabled it. The vacuum bake ovens isolated from the roughing pumps at the loss of power but the Turbo pumps all spun down. As such, the bake load contents were exposed to a Torr*L or two of turbo exhaust but at least reversed viscous flow was prevented. Both ovens remained above 100C and the parts should clean up with the Turbos now restarted. The RGAs for the VBOs were isolated at the time and should not be affected by the spun down turbos. 1100 hrs, local -> Kyle leaving site now. Robert S. and Richard M. still on site.