The dust monitor alarm levels throughout the site have been set to their normal levels. In some cases these levels may need adjusting depending upon what is being done in the space they are monitoring. Please get in touch with Jeff Bartlett before adjusting the dust monitor alarm levels or with any contamination control concerns or questions.
ISO-Clean | 0.3µ Minor | 0.3µ Major | 0.5µ Minor | 0.5µ Major |
---|---|---|---|---|
100 | 200 | 300 | 70 | 100 |
1000 | 2000 | 3000 | 700 | 1000 |
10000 | 20000 | 30000 | 7000 | 10000 |
100000 | 200000 | 300000 | 70000 | 100000 |