EX timing was reporting error with 4th Duotone slave, and h1iopiscex was in error. I found that the timing slave on h1iscex's IO Chassis was still trying to sync up (flashing LED). This system has a separate DC power supply for the timing slave, part of a past noise study. I power cycled both the timing slave and the IO Chassis, and now the code is running correctly and the timing error has been resolved.
started h1pemmy system while at MY. Powered up h1digivideo[0,1,2] and h1auxscript0
started h1pemmx system.
DAQ EDCU is now only complaining about corner station HEPI pump controller (Hugh is on this) and the end station HWS (TiVo is looking into this)
Hugh and Thomas have worked their magic, the EDCU is happy again.