Attached are plots of the injection locking error signal and the servo transfer function. Note that the servo transfer function measurement does not include the SR560 low frequency boost stage that was added before the O2 run. This measurement was taken with - servo gain = 1.00 V - reference level = -1.30 V - error point offset = 0.20 V which are the same ones in use for yonks. The transfer function is comparable to the measurement taken from before the observing run. As noted from yesterday, the RF summing box was placed on to an insulating plastic bag to isolate it from the front end laser breadboard (and hence the table). Touching the box did not appear to affect the injection locking. However I did notice that the frequency servo went into oscillation whenever I touched the box and table. Suspected grounding problem. Probably not noticed, or paid attention to, before because we rarely touch the RF summer. Will be chasing this up over the next few weeks.