Measured the output power of the NPRO with an Ophir 20C-SH power meter (S/N 171175) to be 1.82 W. This was measured before the phase correcting EOM. 1.82 W corresponds to 1.097 V on photodiode (-1797 counts) 0 W corresponds to 0.068 V on photodiode (-111 counts) Pulling up the filter MEDM screen (H1PSLPMC_PWR_NPRO.adl), INMON was ~ -1765 counts. So ~ -32 counts were "missing". I changed the offset to reflect this value. The gain was taken to be -0.00101263150 (ie 1.82 W 1797 counts /). With these changes, the laser MEDM screen displays a value for the output power of the NPRO close to the value measured with a power meter.